Retsch AS 450 basic Manuel d'utilisateur Page 24

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 40
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 23
Operating the machine
24
The main switch (H) is on the left side of the AS450.
Switch the main switch on.
on lights up in the memory display.
The last set vibration amplitude, e.g. 1.00. lights up in the amplitude display
The interval off LED lights up.
Two bars (--)or the last set sieve time light up in the time display.
The AS450 is now ready for operation without interval and for continuous operation
with a vibration amplitude of 1.0 mm.
Pos: 9.17 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 Start en, Unterbrech en, Stoppen @ 0\mod_122 9417133049_ 9.doc @ 5070 @ @ 1
5.8 Starting, Interrupting, Stopping
Pos: 9.18 /00010 Bedienungsa nleitungen Kapitels ammlungen/ AS450_basic/00 07 Bedienung/A S450 basic Mo dul Starten U nterbrechen Stoppe n @ 4\ mod_1329464707 916_9.doc @ 26631 @ 333 @ 1
5.8.1 Start
Press the START/STOP button (B5).
The START/STOP button initiates the operation of the sieve shaker and starts the
countdown for the process duration.
When the START/STOP button is pressed once, the timer starts counting down
and sieving is started.
The decimal point pulsates at intervals of one second, and the digits are counted
down every 6 seconds (or one tenth of a minute).
Display in 1/10 of
a minute
0
1
2
3
4
5
6
7
8
9
Resultant value in
seconds
0
6
12
18
24
30
36
42
48
54
B1
Ab
As
At
At
B2
B3
B4
B5
B6
Vue de la page 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 39 40