Operating the machine
24
The main switch (H) is on the left side of the AS450.
• Switch the main switch on.
– on lights up in the memory display.
– The last set vibration amplitude, e.g. 1.00. lights up in the amplitude display
– The interval off LED lights up.
– Two bars (--)or the last set sieve time light up in the time display.
The AS450 is now ready for operation without interval and for continuous operation
with a vibration amplitude of 1.0 mm.
Pos: 9.17 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 Start en, Unterbrech en, Stoppen @ 0\mod_122 9417133049_ 9.doc @ 5070 @ @ 1
5.8 Starting, Interrupting, Stopping
Pos: 9.18 /00010 Bedienungsa nleitungen Kapitels ammlungen/ AS450_basic/00 07 Bedienung/A S450 basic Mo dul Starten U nterbrechen Stoppe n @ 4\ mod_1329464707 916_9.doc @ 26631 @ 333 @ 1
5.8.1 Start
• Press the START/STOP button (B5).
The START/STOP button initiates the operation of the sieve shaker and starts the
countdown for the process duration.
When the START/STOP button is pressed once, the timer starts counting down
and sieving is started.
The decimal point pulsates at intervals of one second, and the digits are counted
down every 6 seconds (or one tenth of a minute).
Display in 1/10 of
a minute
Resultant value in
seconds